IM 701210-07E
44
MEASURE (GO/NO-GO)
• The following figure shows the menu when the mode is set to ON (automated measurement of waveform parameters).
Turn ON/OFF the automated
measurement of waveform parameters
Section 11.6
Select ON when performing automated
measurement of waveform parameters.
Measurement parameters
Select the measurement parameter to be
automatically measured and the target
waveform.
Delay
Enter settings related to the delay measurement
between channels.
Turn ON/OFF 1 cycle mode
Select whether to compute items related to the
voltage axis or area within the period
determined in measurements such as Rms and
Avg.
Measurement range
Set the time axis range for performing the
automated measurement of waveform
parameters.
Voltage-Axis Parameters
Time-Axis Parameters
Display Example of Automated Measurement
Note
Target waveform for distal, mesial, and
proximal
Select the waveform used to determine the
values of distal, mesial, and proximal that act as
references to the automated measurement of
waveform parameters.
Distal, mesial, and proximal unit
Select % or voltage/temperature.
Distal value
Set the distal value in terms of a percentage or
voltage/temperature.
Mesial value
Set the mesial value in terms of a percentage or
voltage/temperature.
Proximal value
Set the proximal value in terms of a percentage
or voltage/temperature.
High/Low level
Select the method of setting the high and low
levels. Select Auto which sets the level of higher
amplitude to high and the level of lower
amplitude to low or MAX-MIN which sets the
maximum value to high and minimum value to
low.
P-P
Max
Min
High
Low
+OvrShoot
–OvrShoot
+Width
–Width
Period
Distal (90%)
Mesial (50%)
Proximal (10%)
Rise
Fall
High (100%)
Low (0%)
Setup Menu Items
Summary of Contents for ScopeCorder DL750P
Page 1: ...ScopeCorder IM 701210 07E 3rd Edition...
Page 60: ...IM 701210 07E 59 MEMO...
Page 61: ...IM 701210 07E 60 MEMO...