IM 701210-07E
16
Operating the DL750
Operating the DL750/DL750P
/DL750P
Operating the DL750/DL750P
The operation explained in this section is used to observe a signal (probe compensation signal) that is generated by the DL750/DL750P.
Therefore, there is no need to prepare a separate signal generator. Additionally, this operation guide describes an example in which
waveforms are observed using the High-Speed 10 MS/s, 12-Bit Isolation Module (model: 701250, abbreviated name: HS10M12). For
operation using other input modules, see the user’s manual.
Making Preparations before Observing Waveforms
Install the Module
Section 3.3, “Installing Input Modules” in the User’s Manual Part 1
To use the DL750/DL750P in a safe manner, read the warnings given in section 3.3, “Installing
Input Modules” in the user’s manual before installing the module.
Install the module in the input module installation slot on the right side panel.
Module
Right side panel
Attach a cover plate in slots in which
no modules are installed.
Connect the Power Supply
Section 3.4, “Connecting the Power Supply” in the User’s Manual Part 1
To use the DL750/DL750P in a safe manner, read the warnings given in section 3.4,
“Connecting the Power Supply” in the user’s manual before connecting the power supply.
3-pin outlet
Power cord
Power connector
Left side panel
Rated supply voltage:
100 to 120 VAC/200 to 240 VAC
Permitted supply voltage range:
90 to 132 VAC/180 to 264 VAC
Rated supply voltage frequency:
50/60 Hz
Permitted supply voltage frequency range:
48 to 63 Hz