IM 701210-07E
30
Setup Menu Items
CH1 to CH16
Turn ON/OFF the channel
Section 5.1
Input coupling
Section 5.5
To measure temperature, select TC. To
measure voltage (see the previous page), select
DC. To check the ground level, select GND.
Thermocouple type
Section 5.16
Select the type of thermocouple to be used
from K, E, J, T, L, U, R, S, B, N, W, and
Au7Fe.
Bandwidth
Section 5.7
Select the high frequency band to be removed
from the input signal.
Upper and lower limits of the display range
Section 5.9
Set the upper and lower limits of the display
range when displaying measured waveforms
on the screen. The selectable range is –5432
to 5432 [
°
C, K, or
°
F].
Thermocouple Type and Measurement Range
The following 12 types of thermocouples are available.
• K: –200 to 1300 [
°
C]
• E: –200 to 800 [
°
C]
• J: –200 to 1100 [
°
C]
• T: –200 to 400 [
°
C]
• L: –200 to 900 [
°
C]
• U: –200 to 400 [
°
C]
• R: 0 to 1700 [
°
C]
• S: 0 to 1700 [
°
C]
• B: 400 to 1800 [
°
C]
• N: 0 to 1300 [
°
C]
• W: 0 to 2300 [
°
C]
• Au7Fe: 0 to 300 [K]
Note
Waveform label
Section 8.10
You can arbitrary set the waveform label using
up to eight characters.
Unit
Section 5.16
You can set the temperature unit to
°
C, K, or
°
F.
RJC
Section 5.16
Select whether to perform reference junction
compensation by the internal RJC circuit
(ON/OFF).
Burnout
Section 5.16
Set the behavior when the thermocouple input
detects a burnout. Select whether to fix the
measured value at the upper limit of the
measurement range of the respective
thermocouple when a burnout is detected (ON)
or not detect a burnout (OFF).
to
Note
Reference Junction Compensation (RJC)
The DL750/DL750P normally performs reference
junction compensation with the built-in RJC circuit
when measuring temperature with the thermocouple.
When checking the temperature measurement value, or
when using an external reference junction (0
°
C), the
internal reference junction compensation needs to be
disabled (RJC setting must be turned OFF).
• The following figure shows the menu that appears when measuring temperature on channels with the Universal (Voltage/Temp.)
Modules (MODEL: 701261, abbreviated name: UNIVERSAL), Universal (Voltage/Temp.) Modules (with AAF) (MODEL:
701262, abbreviated name: UNIVERSAL (AAF)), or Temperature, High Precision Voltage Isolation Modules (MODEL: 701265,
abbreviated name: TEMP/HPV) installed.