5-23
IM 701730-17E
Commands
5
5.7
COMMunicate Group
The commands in the COMMunicate group are used to make settings and queries aout communications. There
is no front panel key with this function.
:COMMunicate
;
:
?
OPSE
HEADer
<Space>
OFF
ON
<NRf>
<Space>
<Register>
?
OPSR
?
OVERlap
<Space>
<Register>
?
VERBose
<Space>
OFF
ON
<NRf>
WAIT
?
<Space>
<Register>
?
STATus
?
?
LOCKout
<Space>
OFF
ON
<NRf>
?
REMote
<Space>
OFF
ON
<NRf>
?
:COMMunicate:LOCKout
Function
Sets/releases the local lockout.
Syntax
:COMMunicate:LOCKout {<Boolean>}
:COMMunicate:LOCKout?
Example
:COMMUNICATE:LOCKOUT ON
:COMMUNICATE:LOCKOUT? ->
:COMMUNICATE:LOCKOUT 1
Description
This is an exclusive commandfor the USB
interface or Ethernet interface (option).
5.7 COMMunicate Group
:COMMunicate?
Function
Queries all communication settings.
Syntax
:COMMunicate?
Example
:COMMUNICATE? -> :COMMUNICATE:
HEADER 1;OPSE 352;OVERLAP 352;
VERBOSE 1
:COMMunicate:HEADer
Function
Determines whether a header is to be added
(for example:
CHANNEL1:VOLTAGE:PROBE
10
) or nt (example:
10
) when sending a
response to a query; or queries the current
setting.
Syntax
:COMMunicate:HEADer {<Boolean>}
:COMMunicate:HEADer?
Example
:COMMUNICATE:HEADER ON
:COMMUNICATE:HEADER? ->
:COMMUNICATE:HEADER 1