5-110
IM 701730-17E
:TRIGger:WIDTh:CONDition
Function
Sets the trigger condition when using the pulse
width trigger or queries the current setting.
Syntax
:TRIGger:WIDTh:CONDition
{TRUE|FALSe}
:TRIGger:WIDTh:CONDition?
Example
:TRIGGER:WIDTH:CONDITION TRUE
:TRIGGER:WIDTH:CONDITION? ->
:TRIGGER:WIDTH:CONDITION TRUE
:TRIGger:WIDTh:TIME<x>
Function
Sets the pulse width when using the pulse width
trigger or queries the current setting.
Syntax
:TRIGger:WIDTh:TIME<x> <time>
:TRIGger:WIDTh:TIME<x>?
<x>=1 or 2
<time>= 1 ns to 1 s
See the User’s Manual.
Example
:TRIGGER:WIDTH:TIME1 10NS
:TRIGGER:WIDTH:TIME1? ->
:TRIGGER:WIDTH:TIME1 10.0E-09
:TRIGger:WIDTh:TLINkage
Function
Turns the link ON/OFF when changing the
pulse width for the pulse width trigger or queries
the current setting.
Syntax
:TRIGger:WIDTh:TLINkage {<Boolean>}
:TRIGger:WIDTh:TLINkage?
Example
:TRIGGER:WIDTH:TLINKAGE ON
:TRIGGER:WIDTH:TLINKAGE? ->
:TRIGGER:WIDTH:TLINKAGE 1
:TRIGger:WIDTh:TYPE
Function
Sets the criteria type when using the pulse
width trigger or queries the current setting.
Syntax
:TRIGger:WIDTh:TYPE
{IN|OUT|BETWeen|TIMeout}
:TRIGger:WIDTh:TYPE?
Example
:TRIGGER:WIDTH:TYPE BETWEEN
:TRIGGER:WIDTH:TYPE? ->
:TRIGGER:WIDTH:TYPE BETWEEN
:TRIGger:WIDTh:WINDow
Function
Turns the window ON/OFF when using the
pulse width trigger or queries the current
setting.
Syntax
:TRIGger:WIDTh:WINDow {<Boolean>}
:TRIGger:WIDTh:WINDow?
Example
:TRIGGER:WIDTH:WINDOW ON
:TRIGGER:WIDTH:WINDOW? ->
:TRIGGER:WIDTH:WINDOW 1