5-76
IM 701730-17E
:MEASure:USER<x>?
Function
Queries all settings related to the automated
measurement of waveform parameters of the
Dual Area.
Syntax
:MEASure:USER<x>?
<x>=1 to 4
Example
:MEASURE:USER1? -> :MEASURE:USER1:
AREA1:TRACE 1;ITEM PTOPEAK;:
MEASURE:USER1:AREA2:TRACE 2;
ITEM MAXIMUM;:MEASURE:USER2:AREA1:
TRACE 3;ITEM MINIMUM;:MEASURE:
USER2:AREA2:TRACE 4;ITEM RMS;:
MEASURE:USER3:AREA1:TRACE MATH1;
ITEM PERIOD;:MEASURE:USER3:AREA2:
TRACE MATH2;ITEM AVGPERIOD;:
MEASURE:USER4:AREA1:TRACE CONST1;
ITEM TY1INTEG:MEASURE:USER4:
AREA2:TRACE NONE;ITEM XY2INTEG
:MEASure:USER<x>:AREA<y>?
Function
Queries all settings related to the specified
computation parameter/area of the automated
measurement of waveform parameters of the
Dual Area.
Syntax
:MEASure:USER<x>:AREA<y>?
<x>=1 to 4
<y>=1, 2
Example
:MEASURE:USER1:AREA1? ->
:MEASURE:USER1:AREA1:TRACE 1;
ITEM PTOPEAK
:MEASure:USER<x>:AREA<y>:ITEM
Function
Sets the measurement item of the specified
computation parameter/area of the automated
measurement of waveform parameters of the
Dual Area or queries the current setting.
Syntax
:MEASure:USER<x>:AREA<y>:
ITEM {<Parameter>}
:MEASure:USER<x>:AREA<y>:ITEM?
<x>=1 to 4
<y>=1, 2
<Parameter>=
{AVERage|AVGFreq|
AVGPeriod|BWIDth1|BWIDth2|DUTYcycle|
FALL|FREQuency|HIGH|LOW|MAXimum|
MINimum|NOVershoot|NWIDth|PERiod|
PNUMber|POVershoot|PTOPeak|PWIDth|
RISE|RMS|SDEViation|TY1Integ|TY2Integ|
XY1Integ|XY2Integ}
Example
:MEASURE:USER1:AREA1:ITEM PTOPEAK
:MEASURE:USER1:AREA1:ITEM? ->
:MEASURE:USER1:AREA1:ITEM PTOPEAK
:MEASure:USER<x>:AREA<y>:TRACe
Function
Sets the measurement trace of the specified
computation parameter/area of the automated
measurement of waveform parameters of the
Dual Area.
Syntax
:MEASure:USER<x>:AREA<y>:TRACe
{<NRf>|MATH<z>|NONE|CONSt<w>}
:MEASure:USER<x>:AREA<y>:TRACe?
<x>=1 to 4
<y>=1 or 2
<NRf>=1 to 4 (1 to 2 on the DL1720E)
<z>=1 or 2 (1 on the DL1720E)
<w>=1 to 4
Example
:MEASURE:USER1:AREA1:TRACE 1
:MEASURE:USER1:AREA1:TRACE ->
:MEASURE:USER1:AREA1:TRACE 1
:MEASure:USER<x>:COUNt?
Function
Queries the statistical processing count of the
automated measurement of waveform
parameters of the Dual Area.
Syntax
:MEASure:USER<x>:COUNt?
<x>=1 to 4
Example
:MEASURE:USER1:COUNT? ->
:MEASURE:USER1:COUNT 100
:MEASure:USER<x>:{MAXimum|MEAN|
MINimum|SDEViation}?
Function
Queries the statistical value of the automated
measurement of waveform parameters of the
Dual Area.
Syntax
:MEASure:USER<x>:{MAXimum|MEAN|
MINimum|SDEViation}?
<x>=1 to 4
Example
(The following is an example for the maximum
value.)
:MEASURE:USER1:MAXIMUM? ->
:MEASURE:USER1:MAXIMUM 1.000E+00
Description
If the statistics cannot be determined, “NAN
(Not A Number)” is returned.
:MEASure:USER<x>:OPERation
Function
Sets the computing operation of the automated
measurement of waveform parameters of the
Dual Area or queries the current setting.
Syntax
:MEASure:USER<x>:OPERation {PLUS|
MINus|MULTiple|DIVision}
:MEASure:USER<x>:OPERation?
<x>=1 to 4
Example
:MEASURE:USER1:OPERATION PLUS
:MEASURE:USER1:OPERATION ->
:MEASURE:USER1:OPERATION PLUS
:MEASure:USER<x>:STATe
Function
Turns ON/OFF the automated measurement of
waveform parameters of the Dual Area or
queries the current setting.
Syntax
:MEASure:USER<x>:STATe {<Boolean>}
:MEASure:USER<x>:STATe?
<x>=1 to 4
Example
:MEASURE:USER1:STATE ON
:MEASURE:USER1:STATE ->
:MEASURE:USER1:STATE 1
:MEASure:USER<x>:UNIT
Function
Sets the display unit of the automated
measurement of waveform parameters of the
Dual Area or queries the current setting.
Syntax
:MEASure:USER<x>:UNIT <Character
string>
:MEASure:USER<x>:UNIT?
<x>=1 to 4
<Character string>=Up to 3 characters
Example
:MEASURE:USER1:UNIT "UNIT"
:MEASURE:USER1:UNIT ->
:MEASURE:USER1:UNIT "UNIT"
5.17 MEASure Group