:PRESet?
Function
Queries all settings related to the preset
function.
Syntax
:PRESet?
Example
:PRESET -> :PRESET:ALL:TYPE CMOS5;
USERDEFINE:VDIV 2.00;
LEVEL 2.00;OFFSET 0.0E+00;:
PRESET:ALL:PROBE 10;:PRESET:
SELECTCH ALL;CHANNEL1:TYPE CMOS5;
USERDEFINE:VDIV 2.00;
LEVEL 2.00;OFFSET 0.0E+00;:
PRESET:CHANNEL1:PROBE 10;:PRESET:
CHANNEL2:TYPE CMOS5;USERDEFINE:
VDIV 2.00;
LEVEL 2.00;OFFSET 0.0E+00;:
PRESET:CHANNEL2:PROBE 10;:PRESET:
CHANNEL3:TYPE CMOS5;USERDEFINE:
VDIV 2.00;
LEVEL 2.00;OFFSET 0.0E+00;:
PRESET:CHANNEL3:PROBE 10;:PRESET:
CHANNEL4:TYPE CMOS5;USERDEFINE:
VDIV 2.00;
LEVEL 2.00;OFFSET 0.0E+00;:
PRESET:CHANNEL4:PROBE 10
:PRESet:ALL:PROBe
Function
Sets the preset value of the probe attenuation/
current-to-voltage conversion ratio of all
channels.
Syntax
:PRESet:ALL:PROBe {<NRf>|C10|C100}
<NRf>=1, 10, 100, 1000
Example
:PRESET:ALL:PROBE 10
:PRESet:ALL:TYPE
Function
Sets preset values to all channels.
Syntax
:PRESet:ALL:TYPE {CMOS5|CMOS3|
ECL|USERdefine}
Example
:PRESET:ALL:TYPE USERDEFINE
:PRESet:ALL:USERdefine:LEVel
Function
Sets the preset trigger level of all channels.
Syntax
:PRESet:ALL:USERdefine:LEVel
{<voltage>}
<voltage>=8 div on the screen (0.01 div steps)
Example
:PRESET:ALL:USERDEFINE:LEVEL 0V
Description
When set to the current probe, sets the
<Current> value.
:PRESet:ALL:USERdefine:OFFSet
Function
Sets the preset offset voltage of all channels.
Syntax
:PRESet:ALL:USERdefine:OFFSet
{<voltage>}
<voltage>=See the User’s Manual.
Example
:PRESET:ALL:USERDEFINE:OFFSET 0V
Description
When set to the current probe, sets the
<Current> value.
:PRESet:ALL:USERdefine:VDIV
Function
Sets the preset V/div value of all channels.
Syntax
:PRESet:ALL:USERdefine:VDIV
{<voltage>}
<voltage>=2 mV to 10 V (when the probe
attenuation is 1:1, See the User’s Manual.)
Example
:PRESET:ALL:USERDEFINE:VDIV 2V
Description
When set to the current probe, sets the
<Current> value.
:PRESet:CHANnel<x>?
Function
Queries the preset information of the specidied
channel.
Syntax
:PRESet:CHANnel<x>?
<x>=1 to 4 (1 to 2 on the DL1720E)
Example
:PRESET:CHANNEL1? ->
:PRESET:CHANNEL1:TYPE TTL;
USERDEFINE:VDIV 50.000E+00;
LEVEL 0.0E+00;OFFSET 0.0E+00;:
PRESET:CHANNEL1:PROBE 10
:PRESet:CHANnel<x>:PROBe
Function
Sets the preset value of the probe attenuation/
current-to-voltage conversion ratio of the
specified channel or queries the current setting.
Syntax
:PRESet:CHANnel<x>:PROBe {<NRf>}
:PRESet:CHANnel<x>:PROBe?
<x>=1 to 4 (1 to 2 on the DL1720E)
<NRf>=1, 10, 100, 1000
Example
:PRESET:CHANNEL1:PROBE 1
:PRESET:CHANNEL1:PROBE? ->
:PRESET:CHANNEL1:PROBE 1
:PRESet:CHANnel<x>:TYPE
Function
Sets the specified preset or queries the current
setting
Syntax
:PRESet:CHANnel<x>:TYPE
{ECL|TTL|ECL|USERdefine}
:PRESet:CHANnel<x>:TYPE?
<x>=1 to 4 (1 to 2 on the DL1720E)
Example
:PRESET:CHANNEL1:TYPE USERDEFINE
:PRESET:CHANNEL1:TYPE? ->
:PRESET:CHANNEL1:TYPE USERDEFINE
:PRESet:CHANnel<x>:USERdefine?
Function
Queries all preset values when using user-type
presets.
Syntax
:PRESet:CHANnel<x>:USERdefine?
<x>=1 to 4 (1 to 2 on the DL1720E)
Example
:PRESET:CHANNEL1:USERDEFINE? ->
:PRESET:CHANNEL1:USERDEFINE:
VDIV 50.000E+00;LEVEL 0.0E+00;
OFFSET 0.0E+00
:PRESet:CHANnel<x>:USERdefine:LEVel
Function
Sets the preset value of the user-type trigger
level or queries the current setting.
Syntax
:PRESet:CHANNEL<x>:USERdefine:LEVel
{<voltage>}
:PRESet:CHANNEL<x>:USERdefine:
LEVel?
<x>=1 to 4 (1 to 2 on the DL1720E)
<voltage>=8 div on the screen (0.01 div steps)
Example
:PRESET:CHANNEL1:USERDEFINE:
LEVEL 0V
:PRESET:CHANNEL1:USERDEFINE:
LEVEL? -> :PRESET:CHANNEL1:
USERDEFINE:LEVEL 0.000E+00
Description
When set to the current probe, queries the
<Current> value.