5-94
IM 701730-17E
:STATus:FILTer<x>
Function
Sets/queries the transit filter.
Syntax
:STATus:FILTer<x>
{RISE|FALL|BOTH|NEVer}
:STATus:FILTer<x>?
<x>=1 to 16
Example
:STATUS:FILTER2 RISE
:STATUS:FILTER2? -> :STATUS:FILTER2
RISE
Description
For a description of the event register, refer to
chapter 6.
:STATus:QENable
Function
Selects/queries whether messages other than
errors are stored in the error queue.
Syntax
:STATus:QENable {<Boolean>}
:STATus:QENable?
Example
:STATUS:QENABLE ON
:STATUS:QENABLE? ->
:STATUS:QENABLE 1
5.25 STATus Group/5.26 STOP Group
:STATus:QMESsage
Function
Selects/queries whether message content is
appended to responses to "STATus:ERRor?".
Syntax
:STATus:QMESsage {<Boolean>}
:STATus:QMESsage?
Example
:STATUS:QMESSAGE ON
:STATUS:QMESSAGE? ->
:STATUS:QMESSAGE 1
:STATus:SPOLl? (Serial Poll)
Function
Executes the serial polling.
Syntax
:STATus:SPOLl?
Example
:STATUS:SPOLL? -> :STATUS:SPOLL 0
Description
This is an exclusive command for the USB
interface or the Ethernet interface (option).
5.26 STOP Group
The STOP command is used to stop acquisition. The same function can be performed using START/STOP on
the front panel.
:STOP
:STOP
Function
Stops acquisition.
Syntax
:STOP
Example
:STOP
Description
To start acquisition, use “STARt.”