1-3
IM 701730-17E
Overview of the GP-IB Interface
1
1.3
GP-IB Interface Functions
GP-IB Interface Functions
Listener Function
• Allows you to make the settings which you can make using the panel keys on the
instrument, except for the power ON/OFF and GP-IB communications settings.
• Receives commands from a controller requesting output of setup and waveform data.
• Also receives status report commands.
Talker Function
Outputs setup and waveform data.
Note
The listen-only, talk-only and controller functions are not available on this instrument.
Switching between Remote and Local Mode
When Switched from Local to Remote Mode
Remote mode is activated when a REN (Remote Enable) message is received from a
controller while local mode is active.
• REMOTE is displayed on the upper right of the screen.
• All front panel keys except SHIFT+CLEAR TRACE can no longer be operated any
more.
• Settings entered in local mode are retained.
When Switched from Remote to Local Mode
Pressing SHIFT+CLEAR TRACE in remote mode puts the instrument in local mode.
However, this is not possible if Local Lockout has been set by the controller (page 1-6).
• The REMOTE indicator on the upper right of the screen is turned off.
• All front panel keys are operative.
• Settings entered in remote mode are retained.
Note
The GP-IB interface cannot be used simultaneously with other communications interfaces
(USB, or Ethernet).
1.3 GP-IB Interface Functions