IM 1E10C1-E
11-4
11. PARAMETER LIST
Indicator
Name
R/W
Description
Data Range, Units
BRAIN
Terminal
Decimal
Point
Default
Value
b13
B13
VELOCITY CHK 0 to 32.767 m/s
3
R
b14
B14
FLOW DIR
(00) FORWARD
FORWARD
W
(01) REVERSE
b30
B30
REV.SPAN
1 to 3.0000
0 to 4
1.0000
W
b31
B31
BI DIREC HYS
0 to 10%
2
W
b33
B33
FOR.SPAN2
1 to 3.0000
0 to 4
1.0000
W
b34
B34
AUTO RNG HYS 0 to 15%
10
W
b36
B36
LOW ALARM
–10 to 110%
–10
W
b37
B37
L. ALARM HYS
0 to 15%
5
W
—
B60
SELF CHECK
GOOD
R
to
EEPROM DEFAULT
—
C00
ADJUST
R
C01
C01
ZERO TUNING
(00) INHIBIT
ENABLE
W
(01) ENABLE
C02
C02
MAGFLOW ZERO 0 to
±
99.99
2
0.00
W
—
C60
SELF CHECK
GOOD
R
to
EEPROM DEFAULT
—
D00
DISP SEL
R
d01
D01
DISP SELECT
(00) RATE(%)
RATE(%)
W
(01) RATE
(02) FOR. TOTAL
(03) REV. TOTAL
(04) DIF. TOTAL
(05) RATE(%) / FOR. TTL
(06) RATE / FOR. TTL
(07) RATE / RATE(%)
(08) RATE(%) / REV. TTL
(09) RATE / REV. TTL
(10) RATE(%) / DIF.TTL
(11) RATE / DIF.TTL
d02
D02
FL USER SEL
(00) NOT PROVIDED
NOT
W
(01) PROVIDED
PROVIDED
d03
D03
FL USER SPAN
0 to 3000
0 to 4
100
W
—
D10
FL USER UNIT
ASCII 8
W
characters
—
D60
SELF CHECK
GOOD
R
to
EEPROM DEFAULT
Displays span in m/s.
Selects flow direction.
Sets span for reverse direction.
Sets hysteresis when change in flow
direction.
Sets the second span for forward
direction.
Sets hysteresis when change in Auto 2
ranges.
Sets low limit alarm.
Sets hysteresis for low limit alarm.
Displays self-check result.
Automatic zero adjustment functions.
Inhibits or enables automatic zero
adjustment.
Excecutes automatic zero adjustment
and displays zero correction value.
Displays self-check result.
Display functions.
Selects display items on flowmeter
indicator.
Selects whether “PV” is displayed with
user-defined unit.
Sets value displayed in “PV” at 100%
output when “Fl user sel” is provided.
Sets user-defined unit displayed in
“A20.”
Displays self-check result.