IM 1E10C1-E
9-2
9. MAINTENANCE
(2) Status Output (For status output, alarm output or pulse output )
Call up “H01” (TEST MODE) on the setting mode and touch SET key to make the
data display.
Call up “H03” (STATUS OUT) and touch SET key.
Display
Description
Code
00
01
Normal operation
Test output
Content
×
2
×
2
SET
SET
SET
SET
Set “01” and touch SET key twice.
Select the output status and touch
SET key twice.
Code
00
01
02
Normal operation
Contact output status (ON)
Contact output status (OFF)
Content
9.1.2
Setting for Test Output Using BT200
(1) Current Output (Corresponding to flow rate, totalization display and pulse output value)
DATA
Display
Description
Select “TEST” in “H01: TEST MODE”.
Set the value in % of span in “H02:OUTPUT VALUE”. (The figure
shows a 100% setting.)
Flow rate, totalization display and pulse output functions correspond
to this set value.
PARAM
H01:TEST MODE
TEST
H02:OUTPUT VALUE
100 %
H03:STATUS OUT
NORMAL
DIAG
ESC
Code
NORMAL
TEST
Normal operation
Test output
Content
(2) Status Output (For status output, alarm output or pulse output status)
DATA
Display
Description
Call up “H01: TEST MODE” on the setting mode and select
“TEST”.
Select the status to be output in “H03: STATUS OUT”.
PARAM
H01:TEST MODE
TEST
H02:OUTPUT VALUE
100 %
H03:STATUS OUT
NORMAL
DIAG
ESC
Code
NORMAL
TEST
Normal operation
Test output
Content
Code
NORMAL
CLOSED(ON)
OPEN(OFF)
Normal operation
Contact output status (ON)
Contact output status (OFF)
Content