IM 1E10C1-E
3-4
3. INSTALLATION
Please slope the conduit pipe down, and install a drain
valve at the low end of the vertical pipe.
Please open the drain valve regularly.
Drain valve
F030207.EPS
Figure 3.2.7
Conduit Wiring
3.2.6
Connecting to External Instru-
ments
CAUTION
All the devices to be connected to current
output and pulse output must be conformed to
CSA1010, CSA950, or IEC950.
(1)Analog Signal Output(4 to 20mADC)
ADMAG SE
RECEIVER
CUR+
CUR–
F030208.EPS
+
–
+
–
Load Resistance
max. 600
Ω
Figure 3.2.8
Connection for Analog Singal Output
(2)Pulse Output
IMPORTANT
Please give attention to voltage and polarity in
wiring, because it is transister contact (insula-
tion type.)
• In case of the filtering constant of Electric
Counter is more than the pulse width, it makes
signal decreases and can not be calculated
correctly.
• In case of input impedance of electric counter
is large inductive noise from power supply
bring bad influence to measurement. To
calculate correctly, it is recommended to use
shield cable or to make input impedance small
enough within the limits of pulse output of
flowmeter.
ADMAG SE
PULSE OUT
PLS/ALM+
PLS/ALM–
+
–
Protective diode
30 V DC, 0.2A. max*
*In case of FM Nonincendive type, 0.135A max.
Mechanical counter
ADMAG SE
PULSE OUT
PLS/ALM+
+
–
Universal counter or
Electronic counter
Load
PLS/ALM–
F030209.EPS
Figure 3.2.9
Pulse Output Connection
(3)Alarm or Status Output
IMPORTANT
This is a transistor contact(insulated type) so
attention must be paid to voltage and polarity
when making connections.
This output can not switch an AC load. To do
this, another relay (see the figure below) is
required.
* The alarm output works from "close"(Normal) to
"open"(Alarm).
ADMAG SE
ADMAG SE
PLS/ALM+
PLS/ALM–
PLS/ALM+
PLS/ALM–
Load
Protective diode
This connection cannot be made.
Magnetic
valve
AC power supply
F030210.EPS
External
Power supply
30 V DC, 0.2A. max*
*In case of FM Nonincendive type, 0.135A max.
Figure 3.2.10 Contact Output Connection