7. OPERATION VIA HART COMMUNICATOR
ADMAG SE :
Settting fld dev
output to 20mA
HELP
SAVE
ABORT
OK
11
12
Press OK (F4). The output signal of 100% is output.
Voltmeter reading: 5.210
Enter the reading of the voltmeter (5.210), and press ENTER
(F4).
Voltmeter reading: 5.000
Select “Yes” and press ENTER (F4).
“Returning fld dev to original output” appears.
Press OK (F4).
ADMAG SE :
Scaled output: 5.210
equal readout
device?
1 Yes
2 No
HELP
DEL
ABORT
ENTER
13
ADMAG SE :
Enter meter value
5.000000
5.000000
HELP
DEL
ABORT
ENTER
ADMAG SE :
NOTE-Loop may be
returned to automatic
control
DEL
DEL
ABORT
OK
14
‘5.21’
F4
(ENTER)
F4
(ENTER)
F4
(OK)
F4
(OK)
D/A trim should be executed only at single range mode. If D/A trim is executed at Bi
direction mode, it is feared that the current output becomes 108%.
IMPORTANT