IM 1E10C1-E
7-5
7. OPERATION VIA HART COMMUNICATOR
Key operation
There are two choices to select the desired menu item.
1. Use the
or
key to select the desired item, and then press the
key.
2. Press the number key displayed for the desired item.
• To return to the previous display, press the
key, EXIT (F4), ESC (F3) or
ABORT (F3).
Example: Call up the “Tag” to change the tag number.
Check where “Tag” is located in the menu configuration. Then, call up “Tag” on the
display according to the menu configuration.
Device setup
PV
PV AO
Span
process variables
Diag/Service
Basic setup
Detailed setup
Review
:
Line size
PV Damp
Tag
Descriptor
:
ADMAG SE : FI-100
Online
1 Device setup
2 PV
3 PV AO
4 Span
DEL
SET
ESC
ENTER
1
Display
Operation
STU
1
or
ADMAG SE : FI-100
Device setup
1 Process variables
2 Diag/Service
3 Basic setup
4 Detailed setup
5 Review
DEL
SAVE
HOME
ENTER
2
YZ /
3
×
2
or
Display appears when the HART Communicator is turned
on.
Select “Device setup”.
Select “Basic setup”.
Select “Tag”.
The display for Tag setting appears.
(The default value of “Tag” is blank.)
1
ADMAG SE : FI-100
Tag
FI-100
FI-100
HELP
DEL
ESC
ENTER
ADMAG SE : FI-100
Basic setup
1 PV Unit
2 Span
3 Nominal size
4 PV Damp
5 Tag
HELP
SAVE
HOME
ENTER
3
MNO
5
×
4
or
ADMAG SE : FI-100
Tag
FI-100
FI-100
HELP
DEL
ESC
ENTER
4