The PREM Alarm indicator illuminates red. If the self-test is successful, connect the
Accessories and set the generator controls.
C.
PROGRAMMING THE GENERATOR
Programming of User Programs
: Please follow user programs guide.
(Chapter 3 –subsection F)
D.
PATIENT RETURN ELECTRODE MONITORING (PREM)
PREM electrode is used in monopolar surgery modes. Always ensure that
the contact between Patient skin & Patient Return Electrode is adequate.
Bar below PREM icon indicates the contact adequacy.
Icon
Contact Quality
Excellent
100% Contact with patient.
Safe
Contact with patient is proper and safe.
Moderate
Contact with patient is moderate & viable. If
possible try to improve the contact, connect
PREM dual area plate properly. Take appropriate
measures to improve the contact quality.
Poor
It‟s not recommended to continue with the
surgery; Check PREM connection or replace the
notPREM electrodes.
No contact
PREM patient plate is not connected or is
damaged.