True RMS voltmeter (Fluke 8920 or equivalent)
Bipolar and monopolar footswitches and handswitching accessories
Leakage setup per IEC 60601-2-2 clause 19.101 or 19.102
Checking Monopolar High Frequency Leakage Current
1. Connect the 200 Ω load from the active accessory through the current transformer to the
equipotential ground plug on the rear of the generator.
2. Connect the current transformer to a true RMS voltmeter.
3. Connect a monopolar footswitch to the Monopolar 1 footswitch socket at the rear of the
generator.
4. Activate the footswitch in each monopolar mode at the maximum control setting. Record the
leakage current. It should not exceed 150 mA for any mode.
5. If the high frequency leakage exceeds 150 mA, call the Whittemore Enterprises Inc. Service
Centre for further instructions.
Checking Bipolar High Frequency Leakage Current
1. Remove the monopolar accessories, and connect the 200 Ω load from one side of the bipolar
output through the current transformer to the equipotential ground lug on the rear of the
generator.
2. Connect the current transformer to the true RMS voltmeter.
3. Connect a bipolar footswitch to bipolar footswitch socket at the rear of the generator.
4. Activate the footswitch in each mode at maximum control setting. Record the leakage current.
It should not exceed 60 mA for any mode.
5. If the high frequency leakage exceeds 60 mA, call the
Whittemore Enterprises
Inc.
Service
Centre for further instructions.
Summary of Contents for EAGLE TWO PLUS
Page 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Page 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Page 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Page 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Page 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...