USER GUIDE |
EAGLE TWO+
Version: 3.3
51
Single Paddle Footswitch in Sealer/Bipolar Footswitch socket.
-
After Turning on the generator, In Electro Surgery mode single paddle footswitch is
configured for bipolar activations; by default.
-
If user intends to use single paddle foot switch for the sealer activations, it must be
configured.
-
Pressing touch icon in the sealer mode will remove the cross mark from the icon, which
makes the single paddle footswitch functional for sealer operations.
-
For making it functional for bipolar again, user
must configure footswitch into bipolar mode.
By going into bipolar mode and removing
cross mark across the blue footswitch (which is the icon of single paddle footswitch) will make it
functional for bipolar mode.
Double Paddle Footswitch in Monopolar-2/ Bipolar Footswitch socket.
-
In Eagle 2+ double footswitch paddle is configured for Sealer - Bipolar; by default.
-
Footswitch-1 (Yellow in color) can be used for Sealer activations.
-
Footswitch-2 (Blue in color) can be used for Bipolar activations.
3.
Vessel Sealer Mode in Eagle 2+
Double
Paddle
Selection
Single
Paddle
Selection
Activat
ed
Disabled