USER GUIDE |
EAGLE TWO+
Version: 3.3
84
Figure 1
Calibration Step1 - PREM Calibration
Equipment:
PREM plug (modified)
Resistor substitution box
Procedure:
1) Select mode „1:1 REM min‟.
2) Connect PREM plug in the power socket and connect appropriate substitution resistor as
indicated by „Load R‟ on the screen, across the PREM socket.
3) Give single paddle for at least 3-seconds activation to get Fire count reading, once reading is
displayed on the screen press „Write counts to program memory‟ to store the acquired count.
4) Once the memory write operation is successful press Back touch key.
5) Follow the same Steps (1 to 4) for „1:2 REM Max‟
Please Note: Load R is different for 1:2 mode; it will be displayed on the screen
.