USER GUIDE |
EAGLE TWO+
Version: 3.3
80
Checking the Monopolar Cut Output
Check the output for the cut modes.
1.
Verify that the generator successfully completes the self-test.
2.
Connect the test equipment for monopolar output.
a.
Connect one test cable to the left jack in the Monopolar 1/UEM socket. Pass the test cable
through the current transformer and connect the current transformer to the voltmeter.
b.
Use a test cable to short the two pins on the Patient (return electrode) socket.
c.
Connect the second test cable from the voltmeter to both pins of the Patient (return
electrode) socket.
d.
Connect the 300 Ω resistor across the output jacks at the end of the test cables.
e.
Connect the monopolar footswitch to the Monopolar 1 Footswitch socket on the rear panel
of the generator.
3.
Select the Pure mode.
4.
Press the cut up/down keys to set the cut power to 80 watts.
5.
Test the monopolar cut output.
a.
Press the footswitch cut pedal and, while activating the generator, note the output on the
voltmeter.
b.
Release the footswitch pedal.
c.
Based on the voltmeter setting and the current transformer you are using calculate and
record the output current.
6.
Select the Low cut mode and repeat step 5.
7.
Select the Blend1, Blend2 & Hydro mode and repeat step 5. Follow this procedure for every
mode.
8.
Verify that the generator output for each mode is 516
26mA rms.
9.
If the output is outside the specified range, calibrate the monopolar output as described in calibration
steps 4. 2. Then repeat this procedure. If the output for one or more cut modes remains outside
the specified range, call the
Whittemore Enterprises
Inc.
Service Center.
Checking the Monopolar Coag Output
Check the output for the coag modes.
1. Disconnect the 300 Ω resistor and replace it with the 500 Ω resistor.
2. Select the Desiccate mode.
3. Press the coag Up/Down keys to set the coag power to 80 watts.
4. Test the monopolar coag outputs.
a.
Press the footswitch coag pedal and, while activating the generator, note the output on the
voltmeter.
b.
Release the footswitch pedal.
c.
Based on the voltmeter setting and the current transformer you are using calculate and
record the output current.
5. Select Press the Fulgurate mode and repeat step 4.
6. Select Press the Swift, Spray, Hi-Spray mode and repeat step 4. Follow this procedure for
every mode.
7. Verify that the generator output for each mode is 400
20 mA rms. If the output is outside the
specified range, calibrate the monopolar output as described in calibration steps 5.Then repeat
this procedure. If the output for one or more coag modes remains outside the specified range,
call the
Whittemore Enterprises
Inc.
Service Center.
Summary of Contents for EAGLE TWO PLUS