USER GUIDE |
EAGLE TWO+
Version: 3.3
65
a)
Remove all gross matter (blood, mucus, tissue etc.) by wiping the PK Connector
Prepare the connector cable for steam sterilization according to the following cleaning
procedure:
Remove all the residues of blood, tissue and other objects by wiping (do not use abrasive
material) the Connector Cable with a cloth in a mild cleaning solution or detergent capable of
removing organic deposits. Do not use reprocessing solutions.
Rinse thoroughly in running water.
Remove any residues further by cleaning agents with a water damp cloth.
Dry the device thoroughly before sterilizing. Sterilize the connector cable according to the
sterilization procedure.
Summary of Contents for EAGLE TWO PLUS
Page 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Page 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Page 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Page 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Page 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...