Chapter 5.
Programming the Generator
. No footswitch is required in this mode hence provide
flexibility in using bipolar energy.
BiCoag
Alarm:
The system is delivering an alarm after completion of coagulation in endoscopic
& open surgery. This feature is very effective in endoscopic surgeries where surgeon do not have
direct contact with tissue being coagulated. BiCoag Alarm avoids over coagulation of tissue, hence
reduces tissue sticking to instruments and reduces smoke, charring, necrosis & carbonization of
tissue.
B. BLOCK DIAGRAM & DESCRIPTION
System is divided in three basic parts:
1
Processor and control circuit.
2
Variable voltage power supply.
3
Radio frequency output amplifier.
Processor & Control circuit:
This is the main control system that calculates all the system parameters
as per the selected mode and monitors the overall activity of the system.
Variable voltage power supply
:
As per the selected mode and set power settings, power supply
delivers the necessary power as an input for the RF stage.
RF output amplifier
:
This converts the controlled dc voltage in Radio frequency as demanded by
surgical applications.