CHAPTER 7
After Surgery
Electric Shock Hazard:
Warning
–
Never remove the cover of instrument.
After Surgery:
Electric Shock Hazard
Warning
-
Always turn off & unplug the generator before cleaning.
Cleaning
Caution
-
Do not clean the generator with abrasive cleaning that could damage the
generator.
A.
PREPARING THE GENERATOR FOR REUSE
Disconnect the Accessories:
1.
Turn off the generator.
2.
Remove the patient return electrode from the patient if used. Disconnect all other
accessories from the front panel.
If disposable accessory, dispose of it according to the procedures.
If the accessory is reusable, clean & re-sterilize it according to the manufacture‟s
instruction.
3.
Disconnect & store footswitch if used.
B. CLEANING THE GENERATOR
1.
Unplug the power cord from the wall outlet.
2.
Thoroughly wipe all surfaces of the generator & power cord with a mild cleaning solution or
disinfectant & a damp cloth. Strictly follow the procedures for cleaning as mentioned in the
manual. Do not allow fluids to enter the chassis. The generator cannot be sterilized.
3.
Use generator only after machine is completely cleaned.
Instrument Cleaning:
Cleaning is required for all instruments with lumens and hollow spaces.
Abrasive materials should not be used on the any parts because they will damage the
instrument‟s outer surface. Instead, use soft-brushes & cottons clothes.
Use distilled de-mineralized water for the final rinse.
Hard water (high mineral contains) should not be used, it may impact the performance of the
instrument.
Use Neutral PH enzymatic detergent for cleaning whenever possible. Alkaline detergents, if
used, must be completely rinsed from the devices. Do not use corrosive fluids.
Do not exceed 130°C during the washing and sterilization process. Cold soak sterilization is
not typically recommended and, as is always necessary for all instrumentation.
After disassembly, the following manual cleaning steps are important:
1.
All components should be soaked in a blood-dissolving enzymatic solution for at least five
minutes with gentle agitation.
Summary of Contents for EAGLE TWO PLUS
Page 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Page 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Page 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Page 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Page 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...