Caution
Do not stack equipment upon EAGLE TWO+or place the generator on top of other
electrical equipment. This configuration does not allow proper cooling.
Caution
Do not place any
chemical or solution in operating room, may spoil safety of electronic circuit by liquid spillage.
Electromagnetic compatibility (IEC 60601-1-2)
The Shalya fulfills the specifications regarding electromagnetic compatibility.
Voltage Transients (Emergency Generator Mains Transfer)
The EAGLE TWO+operates in a safe manner when the transfer is made between line AC and an
emergency generator voltage source.
Defibrillator Proof
The EA
GLE TWO+
meets specifications for “defibrillator proof” design.
Dimensions and Weight
Length: 42.5 cm
Width: 35.5 cm
Height: 15.5 cm
Weight: < 7.5 kg
Transport and Storage
Temperature range: - 40° to 70°C
Relative humidity: 10% to 100%, condensing.
Operating Parameters
Temperature range: 10°C to 40°C
Relative humidity: 30% to 75%, noncondensing.
Warm-up time
Allow one hour for the generator to reach room temperature before use, if transported or stored
at temperature outside the operating temperature range.
Internal Memory
EEPROM store all the user data & settings.
Audio Volume
Activation Ton
e:
Volume (adjustable): 40 to 65 dB
Frequency:
Bipolar/Sealer: 455Hz
Cut: 580Hz
Coag: 865Hz
Alarm Tone:
Volume (not adjustable): 65 dB
Frequency:
865Hz
Pulse:
Two 800msec.coag tone separated by 800msec.of silence
PREM Contact Quality Monitor
o
Measurement frequency: 80 kHz
10 kHz
o
Measurement current: <10µA
o
Acceptable Resistance Range upto 40%
Summary of Contents for EAGLE TWO PLUS
Page 1: ...EAGLE TWO PLUS INSTRUCTION SERVICE MANUAL ...
Page 38: ...USER GUIDE EAGLE TWO Version 3 3 37 ...
Page 39: ...USER GUIDE EAGLE TWO Version 3 3 38 ...
Page 40: ...USER GUIDE EAGLE TWO Version 3 3 39 ...
Page 56: ...USER GUIDE EAGLE TWO Version 3 3 Whittemore Enterprises Inc 55 STEP 5 STEP 8 STEP 6 STEP 7 ...