UNIT DISASSEMBLY
sssss
87
Remove the Acrylic tape (7401-005) holding the cables to the LCD assembly if replacing the associated cable.
Backlight Cable – 25020-079-70
Touch Controller Cable – 25020-088-70
LCD Data Cable – 25018-042-70
Tilt Switch Cable – 25020-083-70
Braided Loom Process – Item 66
(refer to following Heat Shrink Process if the unit utilized item 67)
Reassemble in reverse order, covering the combined cables with the 6.5” piece of braided loom,
which must be bundled smoothly to fir back through the hinge bracket.
LCD Data
Backlight
Tilt Switch
Touch Controller