TROUBLESHOOTING
103
Affected Leads
Problem
Correction
All
High Freq. Noise.
Adjust low pass filter setting to 150 or
40 Hz (see warning); check proximity to
power cables; check AC filter setting
(50 Hz or 60 Hz).
Remove all portable electronic devices
from the vicinity of the patient / ELI380.
Check to see if the patient has an
implanted muscle stimulator.
Ensure patient is relaxed and muscles
are not tense.
WARNING
: When the 40 Hz
filter is used, the frequency response
requirement for diagnostic ECG
equipment cannot be met.
UNABLE TO SAVE ECG
ECG data too noisy to store.
Directory is full
Correct noise and try
acquisition/storage again.
Erase records. Check Delete rule.
Transmission Troubleshooting Chart
LCD Message
Problem
Correction
TRANSMIT FAILED
Unable to transmit ECG.
Ensure site number is valid. Retry.
Verify ELI Link service is running.
Obtain network log files via service
utility “Dump Log Files” to diagnose
network transmission issue.
ERROR-DICOM Not Enabled
A DICOM communication was
attempted, but the unit is not
configured for DICOM.
Configure the unit to DICOM and
reboot.
UNABLE TO SAVE ECG
No available memory.
ECG data too noisy to store.
Transmit or erase records in the
directory.
Correct noise and try
acquisition/storage again.
DHCP FAILURE
The WLAN module failed to get
an address from DHCP.
Contact your Networking/IT to see if
MAC authentication is required. Verify
the request is getting to the DHCP
server.
Contact Welch Allyn Technical Service.
DPAC FAILURE
WLAN failed to initialize.
Contact Welch Allyn Technical Service.
CAN’T CONNECT TO ACCESS
POINT
A link to the access point could
not be established.
Ensure the IP address, security type,
SSID and password are correct.
If problem persists, contact Welch
Allyn Technical Service.