TROUBLESHOOTING
108
Testing patient cable
Problem
Correction
while watching the screen for artifact. If
you see artifact, replace one or all of
the lead wires.
AM12/AM15E
The sweep square is not
going across the screen but
the time is advancing.
Plug the AM12/AM15E into socket
marked ECG.
No tracings on screen or gaps
in the tracings. Screen shows
Leads Off
The cable between the AM12/AM15E
module and the ELI 380 may have a
broken wire. Move the wire around
starting at the AM12/AM15E module
all the way to the plug going into the
ELI 380. If tracing appear the cable is
bad. This cable is a user replaceable
part.
Contact Welch Allyn Technical
Support to verify warranty status, part
number and procedure to replace the
cable.
Screen shows Searching for
WAM
Change setting to use the AM-XX.
Select the gears icon
, select
WAM/AM-XX, Select Switch To AM-
XX, select Done, press home.
WAM
No tracings
– screen displays
Searching for WAM
Make sure you are within range of the
WAM.
Inspect the battery cap for signs of
damage.
Press the power button on the WAM.
Change the battery checking the
battery compartment for signs of
corrosion.
Make sure you have the correct WAM
for the unit.
Pair the WAM. Press the gears icon,
Select WAM/AM-XX, select
WAM Pairing and follow the prompts.
LCD displays WAM low
battery
Change the AA battery in the WAM.