SPECIAL FUNCTIONS
113
Service Settings Calibrate Cue
NOTE: Cue sensor calibration might be required when the cue sensor, writer assembly, or motherboard
have been changed as a result of device servicing. Calibration is required if “Paper Queue Fault”
message appears on the screen.
Important
: Before selecting this function, verify that the cue mark on the paper and the cue sensor on the printer
are offset. This can be done by making sure the paper is not stopped at the perforation line by opening the paper
tray, pulling out ½ sheet and closing the paper tray.
Select
Calibrate Cue
. Select
Done
.
Auto Test
This button tests the ELI 380’s ability to read and write an ECG file to the flash memory in the unit. The status of
this test will be displayed on the Auto Test screen.
Firmware
This button searches a USB flash drive placed in the USB port on the back of the unit for ELI380.bin file and
reprograms the firmware if found.
Config Files
This button searches a USB flash drive placed in the USB port on the back of the unit for ELI380Config.xml file,
and replaces configuration file if found. It will also search for language files and writer font files and replace them
if found.
Options File
This button searches a USB flash drive placed in the USB port on the back of the unit for the *.sno file, and replace
the .sno file if found.
Dump Log Files
This button writes the following log files to a USB flash drive placed in the USB port on the back of the unit
(X:\ELI380\LOG) in a .txt file format.
LogBattery MMYYYY
LogNetwork MMYYYY
Refer to the information at the end of this section of the manual for detailed information on the log file contents.
Dump Records
This button writes a copy of the ECG and order files on a USB flash drive placed in the USB port on the back of the
unit.