CONFORMANCE TESTING
100
ELI 380 Conformance Test Data Record
Unit Serial #:
Power Testing
Battery 1
– Installed?
Yes
No (if No, skip Battery 1 section)
Running on DC power only (disconnect AC power cord)
Battery Voltage
V
State Of Charge
%
Charge Cycles
Verify Charging is No
On Current (Displayed as Avg. Current)
A (max draw = 1.6A)
(Negative reading when on DC)
Battery Health
%
(Recommend replacement when under 70%.)
Running on DC & AC power (connect AC power cord)
Verify AC power LED indicator illuminates (located above power button). Pass / Fail (Circle
One)
Verify AVG Current value gradually changes from negative to positive (the value will remain
at 0 if the battery is fully charged). Pass / Fail (Circle One)
Verify Charging is Yes (if current is flowing to battery). Pass / Fail (Circle One)
Battery 2
– Installed?
Yes
No (if No, skip Battery 2 section)
Running on DC power only (disconnect AC power cord)
Battery Voltage
V
State Of Charge
%
Charge Cycles
Verify Charging is No
On Current (Displayed as Avg. Current)
A (max draw = 1.6A)
(Negative reading when on DC)
Battery Health
%
(Recommend replacement when under 70%.)