SPECIAL FUNCTIONS
115
Battery Info
ELI 380 communicates with the Lithium Ion batteries’ internal monitoring circuitry to acquire the batteries’ status
and display it on the Battery Info screen. Depending on the electrocardiograph, there can be 1 or 2 batteries in the
ELI 380. The Battery Info screen displays:
Voltage
The battery’s present stored voltage.
Avg Current
The amount of current flowing through the battery. When the battery is charging, the battery is drawing
current from the charging source, which produces a positive value. When the battery is
not
charging, this
value will be negative due to the unit drawing the current from the battery.
Relative State of Charge
The present state of charge of the battery given in percent form.
Battery Health
This value represents the ratio of full capacity (the amount of electrical charge capacity the battery holds at
full charge) to design capacity (the amount of electrical charge capacity the battery is designed to hold). If
below 70%, ELI 380 will notify user to replace battery.
Charge Cycles
The amount cycles the battery has been charged.
Charging
Message will display either Yes, No, or Fault. Yes: the battery is charging; No: the battery is not charging;
Fault: there is a problem with the battery.
Enter Serial Number
This button allows the ability to modify the serial number. This screen is password protected.
USB Device
After connecting ELI 380 via a USB device to a computer, this button activates USB device mode, where user can
access ELI 380 files on the computer.
Dump Config File
This button writes configuration file to a USB flash drive placed in the USB port on the back of the unit. These files