TROUBLESHOOTING
107
LCD message
Problem
Correction
Grinding noise
Paper does not advance and
there is a grinding noise
Make sure you are using the correct
Welch Allyn paper.
Open the paper door and remove the
paper to make sure there is no
cardboard at the bottom of the paper.
Inspect the paper tray to see if there is
any broken plastic.
Bring out ½ sheet of paper and close
the door completely. Go into the
Service menu and press Writer Test. If
the paper does not advance and it is
grinding, press on the lower right
corner of the paper door. If the noise
goes away, there is broken plastic
someplace of the gearbox is out of
alignment.
Contact Mortara Technical Support
Paper comes out crooked or
has lines missing
Verify correct Welch Allyn paper.
Clean the print head with an alcohol
wipe and let dry before testing.
Patient Cable Troubleshooting Chart
Testing patient cable
Problem
Correction
Excess artifact
Verify they are using Welch
Allyn electrodes
Pull the clips off the end of all lead
wires, visually inspect the banana plug
for corrosion, push the clips back on the
lead wires. Inspect the metal part of the
clip that comes in contact with the
electrode for contamination or
corrosion. Clean or replace if needed.
Inspect the lead wires for cracks in the
wires.
Unplug the lead wires from the
AM12/AM15E/WAM and push them
back in again.
Attach a simulator and view waveforms
on the screen for artifact. If artifact is
present, test the unit using another
AM12/AM15E/WAM. If artifact is still
present, contact Welch Allyn Technical
Support.
If no artifact on the simulator, turn off
the simulator and gently bend all wires