PREVENTIVE MAINTENANCE
29
ELI 380 Preventive Maintenance Record
Unit Serial #:
Print device configuration (attach to this report).
Perform Visual Inspection of:
Enclosure/Housing
Contamination
Markings and Labeling
Cabling
Integrity of Mechanical Parts
The ELI 380 may have 1 or 2 batteries installed at time of servicing.
Note Battery 1 Date Code
(YYMMDD) or enter N/A if no battery
Note Battery 2 Date Code
(YYMMDD) or enter N/A if no battery
(Recommend replacement every 2 years.)
Power Testing
Battery 1
– Installed?
Yes
No (if No, skip Battery 1 section)
Running on DC power only (disconnect AC power cord)
Battery Voltage
V
State Of Charge
%
Charge Cycles
Verify Charging is No
Pass / Fail (Circle One)
On Current (Displayed as Avg. Current)
A (max draw = 1.6A)
(Negative reading when on DC)
Battery Health
%
(Recommend replacement when under 70%.)
Running on DC & AC power (connect AC power cord)
Verify AC power LED indicator illuminates (located above power button). Pass / Fail (Circle
One)
Verify AVG Current value gradually changes from negative to positive (the value will remain
at 0 if the battery is fully charged). Pass / Fail (Circle One)
Verify Charging is Yes (if current is flowing to battery). Pass / Fail (Circle One)