UNIT DISASSEMBLY
46
Remove screws (Item 37) from the 7 locations identified.
IMPORTANT:
When replacing a motherboard the following items will need to be accounted for:
Software Version -
If the software version on the new motherboard does not match the software version being
used by the customer, refer to MIS-18-189-01 for instructions on software compatibility and upgrading.
Optional Features:
If any of the following optional features (WLAN, DICOM/Enterprise, Security, Late Potential,
PDQ) need to be enabled on the unit contact Technical Support with the device serial number to receive an “Option
Code” to enable the features. Refer to MIS-18-189-01 for instructions on using the option file to enable features.
Unit Serial Number
– The replacement motherboard will be shipped from the factory with a “generic” service
serial number, which can be changed to the correct serial number by utilizing the service tool located in the
Advanced Menu, Service Settings feature set.
Cue Sensor Calibration –
The cue sensor will need to be recalibrated when replacing a motherboard. The
automatic cue sensor calibration utility is located in the Advanced Menu, Service Settings feature set.