![Waygate Technologies Krautkramer USM Go Operating Manual Download Page 204](http://html.mh-extra.com/html/waygate-technologies/krautkramer-usm-go/krautkramer-usm-go_operating-manual_878369204.webp)
5 Operation
Evaluation according to the DGS method
5-128
Rev. 10 (06/2021)
USM Go
Sound attenuation and transfer correction
There are two possibilities for setting the sound attenu-
ation in the test object:
● before the DGS calibration
in the DGS function group
REF CORR
using the func-
tion
REF ATTEN
● at any time (even after the DGS calibration)
in the function group
MAT ATTN
using the function
TEST ATTEN
The transfer corrections can be set as follows:
● before the DGS calibration
in the DGS function group
REF CORR
using the func-
tion
AMPL CORRECT
● at any time (even after the DGS calibration)
in the function group
MAT ATTN
using the function
TRANSFER CORR.
The settings of the functions
AMPL CORRECT
and
TRANSFER CORR.
have an additive effect, likewise
the settings of the functions
REF ATTEN
and
TEST AT-
TEN
.
Using multiple DGS curves
After turning the DGS evaluation on, at least one curve
is displayed for a certain ERS (equivalent reflector size).
For some test specifications according to DGS, certain
tolerance limits in dB below and/or above this curve
must be monitored.
You can set up to four additional curves by adjusting
their offsets from the original curve in dB values. These
curves have no effect on the displayed measurement
reading or on other settings.
Summary of Contents for Krautkramer USM Go
Page 1: ...Krautkrämer USM Go USM Go Operating Manual Id No 49 155 Rev 10 06 2021 ...
Page 5: ...USM Go Rev 10 06 2021 0 5 Second operating level ...
Page 6: ...0 6 Rev 10 06 2021 USM Go Second operating level continued ...
Page 10: ...0 10 Rev 10 06 2021 USM Go ...
Page 20: ...0 10 Rev 10 06 2021 USM Go ...
Page 21: ...USM Go Rev 10 06 2021 1 1 Introduction 1 ...
Page 35: ...USM Go Rev 10 06 2021 2 1 Standard package and accessories 2 ...
Page 41: ...USM Go Rev 10 06 2021 3 1 Initial start up 3 ...
Page 53: ...USM Go Rev 10 06 2021 4 1 Principles of operation 4 ...
Page 76: ...4 Principles of operation Saving the settings 4 24 Rev 10 06 2021 USM Go ...
Page 77: ...USM Go Rev 10 06 2021 5 1 Operation 5 ...
Page 210: ...5 Operation Evaluation according to the DGS method 5 134 Rev 10 06 2021 USM Go ...
Page 211: ...USM Go Rev 10 06 2021 6 1 Documentation 6 ...
Page 240: ...6 30 Rev 10 06 2021 USM Go ...
Page 241: ...USM Go Rev 10 06 2021 7 1 Maintenance and care 7 ...
Page 247: ...USM Go Rev 10 06 2021 8 1 Interfaces and Peripherals 8 ...
Page 251: ...USM Go Rev 10 06 2021 9 1 Appendix 9 ...
Page 266: ...9 16 Rev 10 06 2021 USM Go ...
Page 267: ...USM Go Rev 10 06 2021 10 1 Specifications 10 ...
Page 278: ...10 12 Rev 10 06 2021 USM Go ...