5 Operation
Calibrating the USM Go
5-30
Rev. 10 (06/2021)
USM Go
Calibration with straight-beam and angle-
beam probes
Case A: With known material velocity
– Set the known material velocity (function group
RANGE
).
– Couple the probe to the calibration block.
– Set the required display range using the function
RANGE
. The calibration echo must be displayed on
the screen.
– Position the gate on one of the calibration echoes un-
til the sound path of the echo is indicated in the mea-
surement line.
– After this, adjust the setting of the function
PROBE
DELAY
until the correct sound path for the selected
calibration echo is indicated in the measurement line.
Example
You are carrying out the calibration for the calibration
range 100 mm using the calibration standard K1 (thick-
ness 25 mm) which is laid down flat.
– Set
RANGE
to 100 mm.
– Set the known material velocity of 5920 m/s (please
see ISO 10863).
– Set the gate so that it is positioned on the first calibra-
tion echo (from 25 mm).
– Read the sound path in the measurement line. If this
value is not equal to 25 mm, adjust the setting of the
function
PROBE DELAY
until it is at 25 mm.
This completes the calibration of the USM Go to the ma-
terial velocity of 5920 m/s with a calibration range of
100 mm for the probe used.
Summary of Contents for Krautkramer USM Go
Page 1: ...Krautkrämer USM Go USM Go Operating Manual Id No 49 155 Rev 10 06 2021 ...
Page 5: ...USM Go Rev 10 06 2021 0 5 Second operating level ...
Page 6: ...0 6 Rev 10 06 2021 USM Go Second operating level continued ...
Page 10: ...0 10 Rev 10 06 2021 USM Go ...
Page 20: ...0 10 Rev 10 06 2021 USM Go ...
Page 21: ...USM Go Rev 10 06 2021 1 1 Introduction 1 ...
Page 35: ...USM Go Rev 10 06 2021 2 1 Standard package and accessories 2 ...
Page 41: ...USM Go Rev 10 06 2021 3 1 Initial start up 3 ...
Page 53: ...USM Go Rev 10 06 2021 4 1 Principles of operation 4 ...
Page 76: ...4 Principles of operation Saving the settings 4 24 Rev 10 06 2021 USM Go ...
Page 77: ...USM Go Rev 10 06 2021 5 1 Operation 5 ...
Page 210: ...5 Operation Evaluation according to the DGS method 5 134 Rev 10 06 2021 USM Go ...
Page 211: ...USM Go Rev 10 06 2021 6 1 Documentation 6 ...
Page 240: ...6 30 Rev 10 06 2021 USM Go ...
Page 241: ...USM Go Rev 10 06 2021 7 1 Maintenance and care 7 ...
Page 247: ...USM Go Rev 10 06 2021 8 1 Interfaces and Peripherals 8 ...
Page 251: ...USM Go Rev 10 06 2021 9 1 Appendix 9 ...
Page 266: ...9 16 Rev 10 06 2021 USM Go ...
Page 267: ...USM Go Rev 10 06 2021 10 1 Specifications 10 ...
Page 278: ...10 12 Rev 10 06 2021 USM Go ...