![Waygate Technologies Krautkramer USM Go Operating Manual Download Page 119](http://html.mh-extra.com/html/waygate-technologies/krautkramer-usm-go/krautkramer-usm-go_operating-manual_878369119.webp)
Rating of welds (function group AWS D1.1)
5 Operation
USM Go
Rev. 10 (06/2021)
5-43
Note
Make sure that all instrument options for the
specific test are calibrated before starting the
rating according to AWS D1.1.
Remember to peak an echo with an amplitude between
45 % and 55 % screen height. A rating is not possible
with other amplitudes.
– Apply couplant and couple the probe to the reference
standard. Maximize the echo from the 1.5 mm side-
drilled hole.
– Select the function
A-START
and adjust the gate A
on the reference echo.
– Vary the gain so that the reference echo is indicated
at 50 % screen height.
– Switch to the function group
AWS D1.1
.
– Select the function
B REFERENCE
and confirm the
selection to store the reference gain.
– Couple the probe to the test object to evaluate a flaw
echo.
– Select the function
A-START
and adjust the gate A
on the flaw echo.
– Vary the gain so that the flaw echo is indicated at
50 % screen height.
– Switch to the function group
AWS D1.1
.
– Store the current gain by means of the function
A INDICATION
. The current gain is stored.
Summary of Contents for Krautkramer USM Go
Page 1: ...Krautkrämer USM Go USM Go Operating Manual Id No 49 155 Rev 10 06 2021 ...
Page 5: ...USM Go Rev 10 06 2021 0 5 Second operating level ...
Page 6: ...0 6 Rev 10 06 2021 USM Go Second operating level continued ...
Page 10: ...0 10 Rev 10 06 2021 USM Go ...
Page 20: ...0 10 Rev 10 06 2021 USM Go ...
Page 21: ...USM Go Rev 10 06 2021 1 1 Introduction 1 ...
Page 35: ...USM Go Rev 10 06 2021 2 1 Standard package and accessories 2 ...
Page 41: ...USM Go Rev 10 06 2021 3 1 Initial start up 3 ...
Page 53: ...USM Go Rev 10 06 2021 4 1 Principles of operation 4 ...
Page 76: ...4 Principles of operation Saving the settings 4 24 Rev 10 06 2021 USM Go ...
Page 77: ...USM Go Rev 10 06 2021 5 1 Operation 5 ...
Page 210: ...5 Operation Evaluation according to the DGS method 5 134 Rev 10 06 2021 USM Go ...
Page 211: ...USM Go Rev 10 06 2021 6 1 Documentation 6 ...
Page 240: ...6 30 Rev 10 06 2021 USM Go ...
Page 241: ...USM Go Rev 10 06 2021 7 1 Maintenance and care 7 ...
Page 247: ...USM Go Rev 10 06 2021 8 1 Interfaces and Peripherals 8 ...
Page 251: ...USM Go Rev 10 06 2021 9 1 Appendix 9 ...
Page 266: ...9 16 Rev 10 06 2021 USM Go ...
Page 267: ...USM Go Rev 10 06 2021 10 1 Specifications 10 ...
Page 278: ...10 12 Rev 10 06 2021 USM Go ...