![Waygate Technologies Krautkramer USM Go Operating Manual Download Page 192](http://html.mh-extra.com/html/waygate-technologies/krautkramer-usm-go/krautkramer-usm-go_operating-manual_878369192.webp)
5 Operation
Distance-amplitude curve according to JB/T4730 and GB 11345 (CNDAC)
5-116
Rev. 10 (06/2021)
USM Go
Adjusting reference lines
You can change the reference lines for RL and EL at
any time. The dB values always refer to the reference
line during recording. These values can also be positive,
i.e. positioned above the reference.
– Select the function group
LINE
.
– Select the function
RL
and press a function key to in-
dividually change the reference line RL.
– Adjust the reference lines SL and EL in the same
way.
Turning the CNDAC evaluation off
You can turn the CNDAC evaluation off at any time.
– Switch to the first operating level.
– Select the function group
CNDAC
.
– Use the navigation to move
upwards
in order to dis-
play the specific function groups and functions for the
JISDAC.
– Select the function group
SETUP
.
– Select the function
CNDAC MODE
and press a func-
tion key to turn the CNDAC evaluation off (
OFF
). The
DAC curve and the evaluation lines are no longer dis-
played in the A-scan.
Note
The DAC curve is not lost by turning the func-
tion off. By turning it on again, you can use
the function
CNDAC MODE
to go back to the
CNDAC evaluation again without losing any
settings.
Summary of Contents for Krautkramer USM Go
Page 1: ...Krautkrämer USM Go USM Go Operating Manual Id No 49 155 Rev 10 06 2021 ...
Page 5: ...USM Go Rev 10 06 2021 0 5 Second operating level ...
Page 6: ...0 6 Rev 10 06 2021 USM Go Second operating level continued ...
Page 10: ...0 10 Rev 10 06 2021 USM Go ...
Page 20: ...0 10 Rev 10 06 2021 USM Go ...
Page 21: ...USM Go Rev 10 06 2021 1 1 Introduction 1 ...
Page 35: ...USM Go Rev 10 06 2021 2 1 Standard package and accessories 2 ...
Page 41: ...USM Go Rev 10 06 2021 3 1 Initial start up 3 ...
Page 53: ...USM Go Rev 10 06 2021 4 1 Principles of operation 4 ...
Page 76: ...4 Principles of operation Saving the settings 4 24 Rev 10 06 2021 USM Go ...
Page 77: ...USM Go Rev 10 06 2021 5 1 Operation 5 ...
Page 210: ...5 Operation Evaluation according to the DGS method 5 134 Rev 10 06 2021 USM Go ...
Page 211: ...USM Go Rev 10 06 2021 6 1 Documentation 6 ...
Page 240: ...6 30 Rev 10 06 2021 USM Go ...
Page 241: ...USM Go Rev 10 06 2021 7 1 Maintenance and care 7 ...
Page 247: ...USM Go Rev 10 06 2021 8 1 Interfaces and Peripherals 8 ...
Page 251: ...USM Go Rev 10 06 2021 9 1 Appendix 9 ...
Page 266: ...9 16 Rev 10 06 2021 USM Go ...
Page 267: ...USM Go Rev 10 06 2021 10 1 Specifications 10 ...
Page 278: ...10 12 Rev 10 06 2021 USM Go ...