Creating the test methods
4-7
Binary solvent manager instrument parameters (Trapping tab):
Tip:
The instrument Comment field is limited to 500 characters. This ensures
correct spacing when generating reports.
Recommended trapping parameters (with a 5-microliter loop):
•
Flow rate: 15
µ
L/min
•
%A: 99.0, %B: 1.0
•
Sample loading time (duration): 1 minute
•
High pressure limit: 5000 psi
•
The system uses the binary solvent manager as the trapping pump for
one minute of the gradient performance test.
20.00
1.00
15.000
99.0
1.0