1-8
System Overview
Sample consumption guidelines
Sample consumption varies depending on your system configuration and
injection mode.
Sample consumption guidelines
System
Configuration
Injection Mode
Requested
Sample Size
(
µ
L)
Sample
Consumed
(
µ
L)
Maximum
Injection
Volume (
µ
L)
a
Loop: 2
µ
L
Needle: 15
µ
L
Partial loop
1
1
1.9
Partial loop with
needle overfill
Not applicable Not applicable Not applicable
Full loop
Auto
2
4
2
1
×
Overfill
2
2
2
2
×
Overfill
2
4
2
3
×
Overfill
2
6
2
Loop: 5
µ
L
Needle: 15
µ
L
Partial loop
2
2
4.9
Partial loop with
needle overfill
2
6
3.8
Full loop
Auto
5
10
5
1
×
Overfill
5
5
5
2
×
Overfill
5
10
5
3
×
Overfill
5
15
5
Loop: 10
µ
L
Needle: 15
µ
L
Partial loop
5
5
9.5
Partial loop with
needle overfill
5
9
7.5
Full loop
Auto
10
20
10
1
×
Overfill
10
10
10
2
×
Overfill
10
20
10
3
×
Overfill
10
30
10