3-6
Configuring System Software
Configuring events
Besides making the signal connections, you must configure the events in
MassLynx.
To configure events:
1.
In the MassLynx window, click Inlet Method.
2.
Select Tools > Instrument Configuration.
3.
Click Events & Triggering, and then click Next.
4.
Select the check boxes that correspond to the event in and event out
connections you made to the mass spectrometer.
Example:
If you connected the inject start terminals on the sample
manager connector to the number 1 Event I/P connector on the mass
spectrometer, select box 1 in the Event In section.
Choose Events dialog box:
5.
Click Next. In the Configuration Successful window, click Finish.
Summary of Contents for nanoACQUITY UPLC
Page 7: ...vii ...
Page 16: ...xvi ...
Page 20: ...xx Table of Contents ...
Page 32: ...1 12 System Overview ...