Preparing the sample manager
2-15
To prime the sample manager:
1.
In the nanoACQUITY UPLC Console, select Sample Manager from the
system tree.
2.
Click Control > Prime syringes.
Alternative:
Right-click in the MassLynx sample manager control panel,
and then click Prime syringes.
Prime Syringes dialog box:
3.
Select Sample syringe and wash syringes.
4.
Type the number of primes in the “Number of cycles” text box. The
default value is 1.
Recommendation:
Waters recommends 5 to 7 cycles when you are
changing solvents.
5.
Click OK.
Tip:
Each prime takes approximately 2 to 4 minutes.
6.
When the system status is Idle, priming is finished. Click Close.
Washing the sample manager needle
Washing the needle is an optional procedure that flushes strong and/or weak
wash solvent through the needle and injection port. Washing the sample
manager needle removes contaminants from its inner and outer surfaces and
from the external piercing needle and injection port. You can also perform a
needle wash to ascertain proper flow through the waste tubing and to confirm
that the needle wash system is primed and properly operating.