Using the operator interface
3-37
To program a new timed event:
1.
Press the METHOD (Shift A/B) key on the detector keypad. The Method
choice list appears.
Method choice list
2.
Press 1, Timed events. An active field for entering the time of the event
appears.
3.
Enter the time for the event. When you begin entering the time,
additional fields appear.
6
Auto zero
Does not
apply
Does not apply
Yes
7
Lamp
1. Off
2. On
Off
No
8
Switch 1
1. High
2. Low
3. Pulse
4. Rect wave
High
No
9
Switch 2
1. High
2. Low
3. Pulse
4. Rect wave
High
No
10
Threshold
AU
–4.0000 to 4.0000 or
variable, depending on
output selection
Yes
Timed event parameters
Number
Event
Units
Range or Default
Specify
Channel
Summary of Contents for 2489