4-24
Maintaining the Detector
Tip:
Be sure to enter the 9-digit lamp serial number and not the lamp
part number when performing this procedure.
4.
Enter the 9-digit serial number of the new lamp in the active field. This
field accepts numeric entries only.
Change Lamp screen
5.
Press Enter to store the serial number and move to the Date installed
field.
6.
Select the month from the choice list. Press Enter twice to update the
month and move to the date field.
7.
Enter the number for the day of the month the lamp is installed, and
then press Enter to enter the date and move to the Year field.
8.
Enter the year (last two digits only). Press Enter to update the year and
move to the Hours field.
Tip:
The Hours field is optional. If you are re-using a lamp that has
pre-existing hours on it, enter the number of hours the lamp has already
been used, and then press Enter. If this is a new lamp, the hours are 0.
9.
Press the HOME key.
10. At the “OK to store” message, press Enter to store the serial number and
date of installation, or press Cancel to cancel the entry.