3-60
Preparing the Detector
Before you begin
Recommendation:
To ensure accurate results, use a 10-mm path length
quartz cuvette and matched pairs (from the same manufacturing lot) of quartz
cuvettes for your zero and sample scans.
Before beginning a scan using the cuvette:
1.
Flush the flow cell with the eluent you are going to scan.
2.
To ensure a clear and clean cuvette, wipe the clear portion of the cuvette
with low-lint, nonabrasive tissue wipes.
Cuvette scanning procedure
To begin a cuvette scan:
1.
Remove the detector front left panel cover.
2.
To remove the cuvette holder, slide it toward you.
Removing the cuvette holder
TP02808
Summary of Contents for 2489
Page 8: ...viii...
Page 32: ...1 16 Theory and Principles of Operation...
Page 134: ...3 66 Preparing the Detector...
Page 193: ...Warnings that apply to all Waters instruments A 9...
Page 204: ...B 6 Detector Specifications...
Page 220: ...D 14 Solvent Considerations...