Using the operator interface
3-31
Calibration successful message
3.
Press Enter to complete calibration.
A “Calibration complete” message is displayed momentarily. Other
messages such as “Optimizing system performance” and “Restoring last
setup” may appear before the display returns to the absorbance screen.
Result:
If calibration is successful, the error message (<Error>) that was
displayed on the absorbance screen before you recalibrated the detector
disappears.
4.
If calibration is not successful, retry, recycle power to the detector, or see
Operating the detector in single wavelength mode
The detector is optimized for single wavelength operation, which is the default
operating mode.
To invoke single wavelength mode:
1.
From the absorbance (or HOME) screen, press the
λ/λλ
key (Shift Auto
Zero) if the detector is in dual wavelength (the wavelength icon shows
λλ)
mode. The detector displays a message indicating it is switching to
single wavelength operation.
Wavelength mode change message (for single wavelength)
Summary of Contents for 2489
Page 8: ...viii...
Page 32: ...1 16 Theory and Principles of Operation...
Page 134: ...3 66 Preparing the Detector...
Page 193: ...Warnings that apply to all Waters instruments A 9...
Page 204: ...B 6 Detector Specifications...
Page 220: ...D 14 Solvent Considerations...