D-12
Solvent Considerations
particular analysis. Because of the diversity possible within a given sample,
scanning over a range of wavelengths may be necessary to determine the best
wavelength for a particular analysis.
Electronic absorption bands of representative chromophores*
Chromophore
Chemical
Configuration
λ
max
(nm)
∈
max
(L/m/cm)
λ
max
(nm)
∈
max
(L/m/cm)
Ether
—O—
185
1000
Thioether
—S—
194
4600
215
1600
Amine
—NH2
195
2800
Thiol
—SH
195
1400
Disulfide
—S—S—
194
5500
255
400
Bromide
—Br
208
300
Iodide
—I
260
400
Nitrile
—C
≡
N
160
—
Acetylide
—C
≡
C—
175-180
6000
Sulfone
—SO2 —
180
—
Oxime
—NOH
190
5000
Azido
>C=N—
190
5000
Ethylene
—C=C—
190
8000
Ketone
>C=O
195
1000
270-285
18-30
Thioketone
>C=S
205
strong
Esters
—COOR
205
50
Aldehyde
—CHO
210
strong
280-300
11-18
Carboxyl
—COOH
200-210
50-70
Sulfoxide
>S—O
210
1500
Nitro
—NO2
210
strong
Nitrile
—ONO
220-230
1000-2000
300-400
10
Azo
—N=N—
285-400
3-25
Nitroso
—N=O
302
100