Using the operator interface
3-19
Primary and secondary function (method) parameters
Function
Screen
Type
Units
Range
Default
λ
(Wavelength)
1
(Absorbance
screen)
Numeric
nm
Integer 190 nm to
700 nm
254 nm
AUFS
1
Numeric
AUFS
0.0001 to 4.0000
2.0000
Analog rate
2 (of 4) or
2 (of 5)
Choice
Hz
(
λ
): 10, 20, 40, 80
(
λλ
): 1 or 2
10
1
Filter time
constant
2 (of 4) or
2 (of 5)
Numeric
Sec
Slow
Normal
Fast
Other:
(
λ
): 0.0125 to 5.0
(
λλ
): 0.5 to 5.0
0 to disable
filtering
Normal
(for Other:)
1.000
Data out
(single
λ
)
3 (of 4)
Choice
None
Absorbance A
Absorbance
A
Data out (dual
λ
)
3 (of 5)
Choice
None
Absorbance A
Absorbance B
Maxplot A, B
Difference A-B
Ratio A/B
Absorbance
A
Auto zero on
inject
3 (of 4) or
3 (of 5)
Check
box
None
Selected/
Not selected
Checked
Auto zero on
λ
changes
3 (of 4) or
3(of 5)
Choice
None
To baseline
To zero
Disable
To zero
Voltage offset
4 (of 4) or
4 (of 5)
Numeric
mV
Integer –2000 to
2000
0
Chart polarity 4 (of 4) or
4 (of 5)
Choice
None
+
–
+
Minimum AU
5 (of 5)
Numeric
AU
0.0001 to 4.0000
0.1000