Replacing the lamp
4-19
3.
After the lamp is extinguished, allow the lamp to cool for a minimum of
30 minutes.
4.
Remove the front left panel cover, as described in
5.
Disconnect power to the lamp and disconnect the cord.
Lamp assembly and power connector
6.
Loosen the two captive screws in the lamp base.