Periodic Safety Testing
Warning:
Earth leakage current and the ground integrity of the
metal chassis should be tested regularly, at least
once every six months, by trained personnel. If the
system fails either of these tests, discontinue the
use of the equipment immediately and contact
your service provider.
Earth Leakage Current should not exceed 100
microAmperes when measured in compliance with
EN60601-1 and IEC 601-1.
Ground Integrity, when measured at 30 Amperes
should not exceed 0.1 Ohm from the ground pin on
the AC plug, to all parts of the metal chassis, including
drawers, shelves, and camera arm (if provided).
Périodicité des tests de sécurité
Mise en Garde: Au moins tous les six mois, le courant de fuite et
l’intégrité de la mise à la terre du châssis métallique
devraient être testés par un personnel dûment
formé à cet effet. Si l’équipement ne passe pas l’un
ou l’autre de ces tests, son emploi doit être
immédiatement interrompu, et le responsable
chargé de l’entretien doit être contacté.
Pour être conforme aux normes EN60601-1 et IEC-
601-1, le courant de fuite ne doit pas dépasser 100
microampères.
L’intégrité de la mise à la terre, mesurée à 30
ampères, ne doit pas dépasser 0,1 Ohm de la borne
de terre sur la fiche CA, à tous les secteurs du châssis
métallique, à savoir les tiroirs, étagères et bras de
caméra (optionnel).
*
†
•
‡
*
†