Section 6 —Service Testing & Calibration
Page 6-12
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
3. The cuff pressure should increase until approximately 300-315
mmHg. Verify that the
Pass
box for
Adult
contains a checkmark.
Figure 6-14: Adult Test Pass
4. Click on the
Test Infant Mode.
A checkmark should appear in
the box before it. Click
Start
.
Figure 6-15: Test Infant Mode