Section 2 —Service Menus
Page 2-14
VitalCare
TM
506N3 Series
Service Manual
Criticare Systems, Inc.
High Pulse
Select the high alarm limit for pulse rate. Choices are 80 to 250 bpm
and
Off
. Resolution is 1 bpm. The factory default value is 150 for
Adult and Pediatric modes and 180 for Neonate mode.
Low Pulse
Select the low alarm limit for pulse rate. Choices are off and 20 to 150
bpm and
Off
. Resolution is 1 bpm. The factory default value is 40 for
Adult and Pediatric modes and 90 for Neonate mode.
High SpO
2
Select the high alarm limit for SpO
2
. Choices are 70-98%, and
OFF
.
the factory default setting is
OFF
for both adult and neonatal modes.
The resolution is 1% blood oxygen saturation. The factory default
value is
OFF
.
Low SpO
2
Select the low alarm limit for SpO
2
. Choices are
OFF
and 1 to 98%.
The factory default value is 90.
If
LOW SPO2
is set to 98%, the
HIGH SPO2
alarm may not be
changed from the
OFF
setting.
The
LOW SPO2
setting returns to a minimum value of 90% after a
power cycle.
High Systolic
Select the high alarm limit for systolic blood pressure. Choices are 75
to 240 mmHg and
Off
. The factory default value is 200 for Adult and
Pediatric modes and 140 for Neonate mode.
Low Systolic
Select the low alarm limit for systolic blood pressure. Choices are 50
to 150 mmHg and
Off
. The factory default value is 50.
High Diastolic
Select the high alarm limit for diastolic blood pressure. Choices are
50 to 180 mmHg and
Off
. The factory default value is 100 for Adult
and Pediatric modes and 80 for Neonate mode.
Low Diastolic
Select the low alarm limit for diastolic blood pressure. Choices are 15
to 50 mmHg and
Off
. The factory default value is 30 for Hospital
defaults. The factory default value is 40 for Adult and Pediatric modes
and 30 for Neonate mode for Alternate Care defaults.
High MAP
Select the high alarm limit for mean arterial blood pressure. Choices
are 70 to 200 mmHg and
Off
. The factory default value is 150 for
Adult and Pediatric modes and 100 for Neonate mode for Hospital
defaults. The default value for all modes is
Off
for Alternate Care
defaults.
Low MAP
Select the low alarm limit for mean arterial blood pressure. Choices
are 25 to 125 mmHg and
Off
. The factory default value is 50 for Adult
and Pediatric modes and 40 for Neonate mode for hospital defaults.
The default value for all modes is
Off
for Alternate Care defaults.
Summary of Contents for 506DN3
Page 12: ......
Page 56: ......
Page 62: ......
Page 90: ......
Page 156: ......
Page 184: ......
Page 225: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ...DO NOT SCALE PRINT DO NOT SCALE PRINT REV DATE DESCRIPTION BY...
Page 253: ......
Page 254: ......